| **Latest
Updates on TEM and SEM Labs**
Home
Microscope Training
Instructions Camera
Booking
Sample preparation
Methodologies Associated
instrumentation
Scanning
EM
Image Galleries
Background to EM
Contract
Work
Research
Publications
Research example: AD and
the Tauopathies website
Interesting EM and other scientific
links
Compiled by: Julian
Thorpe
| |
Why do we use Electron Microscopes?

General
 |
You are probably familiar with light microscopes,
but what are electron microscopes, why were they developed and why
are they so widely used? |
 |
The answer is all about resolution (the ability to
see fine detail in the image). Resolution is dependent (among other factors)
upon the wavelength of the illumination used in the microscope. |
 |
The Light Microscope uses light as the illuminating
source. The highest effective magnification of a light microscope is about
1,000X; at magnifications beyond this no further detail in the image is
acquired. The best possible theoretical resolution of a light microscope
is about 0.25 microns. |
 |
The electron microscope (or EM), however, has
a beam of electrons as its source of illumination. As the wavelength
of this electron beam is shorter than that of light then resolution is
improved. In 'transmission' EMs (see below) this is typically around
a few tenths of a nanometre (i.e. 0.0002 microns). |
 |
Because of this improved resolution the image in an electron
microscope may be magnified up to around 500,000X and we can see much finer
detail in the specimen. |


Types of Electron Microscope
 |
There are two basic types of EM: the Transmission EM
(TEM) and the Scanning EM (SEM). |
 |
In the TEM the electron beam passes through the specimen,
which is routinely a thin section or a particulate sample (e.g.
proteins or virus particles). |
 |
The TEM is used to observe the fine internal structure (ultrastructure)
of the specimen (i.e. all the cellular organelles). |
 |
In the SEM the electron beam is scanned over
the specimen, which is usually what we call a ‘bulk’ sample (e.g.
a whole insect, plant part, etc.). |
 |
The SEM is therefore utilised to discern the external
appearance (or surface morphology) of the specimen. |
 |
Both types of EM operate under a high vacuum as the
electron beam has limited energy (gas or water molecules would hinder the
passage of the electrons). |

The Scanning Electron Microscope (SEM)
SEM Preparation
 |
The specimen is firstly fixed (i.e. preserved
in as close to its living state as possible). |
 |
Specimens are then dehydrated (i.e. their water is
removed and replaced with ethanol [alcohol]). This is required because
electron microscopes operate under a high vacuum and therefore no water
should be allowed into the microscopes. |
 |
Next, the specimens are ‘critical-point dried’. This
is carried out in a specialised pressure vessel (a critical-point dryer)
and involves replacing the ethanol with liquid carbon dioxide (CO2). The
liquid CO2 is then ‘sublimed’ (converted from liquid to gas) instantaneously
by heating. |
 |
The critical-point drying produces a perfectly dry but turgid
(i.e. not collapsed or shrivelled) specimen. |
 |
The specimen is then stuck onto a special SEM sample holder
(called a stub) and then ‘sputter-coated’ with a very thin
layer of gold in a coating device. This gives the sample a conductive
layer and prevents charging in the SEM. |
 |
The specimen is now ready for viewing in the SEM.
|
What do we use SEMs for?
 |
The SEM is used for looking at the surface features
(external morphology) of a specimen in fine detail. |
 |
Some examples of specimens commonly looked at in our
lab are: Drosophila (fruitflies), feather mites, the cochlea (part of the
inner ear) and floral parts. |
 |
The SEM may also be used to find out which elements
are present within a specimen. This is carried out with an elemental
analysis detector, which can collect the x-rays given off by
a specimen when the electron beam hits it. |
 |
The useful feature of these x-rays is that each individual
element within a sample gives off x-rays of a particular type.
Therefore, if we look at the types of x-rays collected, we can tell which
elements have emitted them. |

The Transmission Electron Microscope (TEM)
 |
The specimen is firstly fixed (i.e. preserved in as
close to its living state as possible). |
 |
Specimens are then dehydrated (i.e. their water is
removed and replaced with ethanol [alcohol]). This is required because
electron microscopes operate under a high vacuum and therefore no
water should be allowed into the microscopes. |
 |
The specimens are then placed in a liquid resin (similar
to an araldite-type glue). |
 |
After a few days (and several resin changes) the specimens
are heat-polymerised (the resin is hardened) to produce a firm specimen
block. |
 |
Very thin sections are then cut on a special section-cutter
called an ultramicrotome. |
 |
The sections are then stained in solutions of heavy
metals (usually uranium and lead) which produce contrast in
the image by causing scattering (diversion) of the electron beam.
|
What do we use TEMs for?
 |
The TEM is used for looking at the fine detail of
structure within the specimen. |
 |
Within cells this fine detail is called ultrastructure. |
 |
Being able to see the ultrastructure within cells means that
we can see all the cellular organelles in detail. |
 |
We can see changes in the ultrastructure within cells (e.g.
after a drug treatment) by a method called morphometric analysis. |
 |
Morphometric analysis is carried out by taking images with
a digital camera. The digital images are then analysed on
a computer and the changes in relative amounts of the organelles
can be measured. |
 |
The TEM may also be used to see where specific proteins
are within a cell. This is done by a method known as immunogold labelling. |

|