School of Life Sciences                                                                 

The Sussex Centre for Advanced Microscopy

**Latest Updates on TEM and SEM Labs**

Home

Microscope Training  Instructions  Camera   Booking      Sample preparation  Methodologies  Associated instrumentation

Scanning EM 

Image Galleries      

Background  to EM

Contract Work

Research

Publications

Research example: AD and the Tauopathies website

Interesting EM and other scientific links

Compiled by: Julian Thorpe

 

Use of the Transmission Electron Microscope (TEM)

**IMPORTANT** Please don't use the below as a substitute for having training with Julian on the TEM, WHICH YOU MUST HAVE BEFORE USING THE MICROSCOPE(!),but you may find this  useful to refresh your memory after training when you are actually sat at the TEM.

Please e-mail Julian or phone 7585 if you would like to arrange a training session with me. 

     Also, if you have any problems with the TEM, and I am not around at the time, please mail me with a description of the problem.

Specimen rod removal ; grid insertion ; specimen rod insertion   
initial viewing of samples
TEM ALIGNMENT: OL aperture centering; voltage centering ; astigmatism correction  
high resolution imaging: changing spot size
magnification calibration
useful features on the TEM: specimen pre-heating ; memorizing specimen positions ; specimen 'trace' ; image rotation  
electron diffraction imaging
TEM troubleshooting
Gatan Ultrascan 1000 CCD Camera: Instructions A
Gatan Ultrascan 1000 CCD Camera: Troubleshooting

Specimen Rod Removal:

The specimen rod is normally left in the TEM (to keep it clean within the vacuum). 

N.B.: During the following, please keep a firm grip of the specimen rod otherwise the vacuum will tend to suck it back into the column and could cause expensive damage!! And if you feel any resistance, please do not use any force(!!) - just come and ask me to help.

To withdraw the specimen rod, 
Pull specimen rod outwards carefully until it reaches a natural stop
Turn to the right (about 45 degrees) and pull out further until it reaches another natural stop
Then turn to the left (again about 45 degrees)
Now flick the switch (below the specimen airlock) to air, wait a few seconds (until you hear a click) 
Withdraw the rod carefully (keeping it in a horizontal plane) from the TEM

Grid Insertion:

N.B.: During the following, please do not touch the rod at all with your hands [below the level of the vacuum O-ring]; grease on the rod will lead to charging of the electron beam and imaging problems
Place the specimen rod on the special rod holder and draw back the securing flange by placing the wooden stick in the hole
Now remove the small plate that secures the grids in place with forceps and place your 2 grids carefully over the grid-holes (N.B. grid position 1 is that nearest the tip of the holder)
Replace the plate over the grids and carefully slide back the securing flange with the stick
Visually check that the grids are still central within the grid holes and invert and gently tap the rod to ensure that they are secure and will not drop out within the TEM!!

Specimen Rod Insertion:

 
Carefully insert the rod into the specimen airlock so that the small protruding metal stud goes into the groove (at 12 o/c) on the outer face of the airlock
Hold in this position and flick the switch to evacuate
N.B.: During the following, please keep a firm grip of the specimen rod otherwise the vacuum will tend to suck it back into the column and could cause expensive damage!! And if you feel any resistance, do not use any force!!!! 
Wait for the green light to illuminate and then carefully turn the rod to the right (45 degrees) and carefully allow the rod to slide in until it reaches a natural stop
Now turn to the left (45 degrees) and carefully allow to move forward completely into the imaging position. 
You can move from grid 1 to 2 by rotating the knob right at the end of the specimen rod (fully one way or the other).

Initial Viewing of Your Sample:

 
If the accelerating voltage is not already on (illuminated), press 'acc voltage' ready/off button (left console of EM) and select the voltage required (usually 100kV). Now turn up the filament current (just below acc voltage button) quite slowly and fully clockwise. The electron beam is now on.
Initial observations are more easily accomplished in 'low mag' mode. Select the 'low mag' function button (upper right portion of left console) and then withdraw the objective aperture from the EM. The objective aperture is at the same level as the specimen rod on the left side of the EM. To withdraw the aperture swing the rod at the base of the aperture rod unit from left to right (please do this carefully as it is on a sprung mechanism).

N.B. Please note that the specimen rod and the objective aperture assembly are the only parts of the EM column you should ever need to touch!!
From hereon, the only controls on the EM console that you need to use are the following: (left console) 'low mag' and 'zoom' function buttons, 'magnification' and 'brightness' control knobs; (right console) focus 'coarse' button, 'focus' and 'brightness centring' control buttons. These are all clearly labelled and easily accessible in the dark when you are familiar with their position. Please avoid touching anything else. If you get into trouble please cry for help rather than attempt things which may result in the need for a major re-alignment of the EM!
Now you should see a very low magnification image of your grid; if you do not then turn the magnification to the lowest mag (fully anticlockwise until you hear the beeps) and then adjust the brightness so that the electron beam fills the screen. Now you can look over your sections and select an area of interest which you should position in the screen centre.
Now you need to carefully replace the objective aperture by swinging the rod to the left and then select the 'zoom' function. If you do not see anything after this, follow this procedure: select the lowest magnification and then turn the brightness control knob so that the electron beam fills the screen. If, after this, you still do not see anything (or at best a rather dark image) then the objective aperture is probably off-centre. Objective aperture centring is achieved by x and y shifts on the aperture rod assembly (these are the smallest knob (furthest from the EM column) and the small knob on the right of the assembly). Please go to OL Aperture Alignment for details as to how to do this.

Routine Imaging:

To achieve a good final image, follow the procedure below:

(i) Lift the central portion of the fluorescent viewing screen (by raising the large flange to the left of the viewing chamber) so that it is perpendicular to the binoculars. Slide the binoculars into position and carefully focus the binoculars themselves onto the small cross in the screen centre.

(ii) Now select the magnification you require and adjust and centre the beam via the brightness and brightness centring controls.

(iii) Focus on your specimen firstly with the focus 'coarse' button selected (illuminated). When at best focus, de-select the 'coarse' button and now you can fine-focus the image.

(iv) At this stage you need to correct for astigmatism in the objective lens. (Astigmatism is present in any lens whether electromagnetic or glass; the main symptom of astigmatism is that, as you go through focus, the image appears rather blurred in one plane when under-focused and in a plane at 180 degrees to this when over-focused. If astigmatism is very bad then it is impossible to focus the image at all well). Please go to Astigmatism Correction for full details as to how to do this.

TEM Alignment

The best possible image will only be achieved when the TEM is aligned properly. The Hitachi engineers are very good and they leave the TEM nicely-aligned, in respect of all the more complex alignment procedures, after every service visit. However, there are a number of basic alignments that vary with kV, spot size, aperture used, etc. and these are worth checking when you start to use the microscope and after changing the spot size. viz.:

the objective aperture should be centred precisely: see detail on this procedure

the voltage centre should be aligned: see detail on this procedure

the astigmatism should be corrected precisely: see detail on this procedure

High Resolution Imaging

N.B. For all the below, please ask myself for help if you have not received training in any particular routine.

There are various factors that will help you get the most out of your sample imaging. Some of these will be influenced by the type of sample you are examining (eg. is it electron beam-sensitive, low or high contrast, relatively dense or thin, etc.). The general rules of thumb are:

Higher accelerating voltages (kVs) will provide better resolution but decreased contrast.

Contrast can be enhanced by using a smaller aperture (3 or 4) but this is at the expense of some resolution.

A smaller spot size will provide better resolution but the electron beam will become dimmer. There are 8 spot sizes, with 1 giving the brightest, lowest resolution image and 8 the dimmest, highest resolution. Spot size 6 is a reasonable compromise for most viewing with 4 being better for higher resolution imaging. Please go to Changing Spot Size for details as to how to do this.

Magnification Calibration:

N.B. If you want to obtain precise measurements of your sample, I now have a calibration sample for higher magnifications on the TEM. This consists of catalase crystals that have known lattice spacings (of 8.75 and 6.85nm). I can show you individually how to do this as and when required.

A couple of important points to remember before taking measurements:

Firstly, turn the magnification knob up to the highest magnification and then down again. (This standardizes the lenses for effects of hysteresis).
Secondly, the specimen Z height affects the magnification. If you don't know where or what this is, then I am quite happy (!) as it routinely doesn't need to be touched. If ever any of you feel the Z height needs checking or changing, please ask me to do that and then I can alert everyone via email that the magnification calibration should be re-checked.
Thirdly, I would suggest that if your measurements need to be absolutely precise, then you should calibrate the magnification before you start using all the conditions (kV, spot size, etc.) and magnifications that you will be using to image your sample..
***Very Important*** Finally, I strongly suggest that you actually make a note of your own calibration numbers and make sure they are correct before you start a new session. I can show you how to check these numbers. In the future I am going to try and sort out a system whereby we can all have our own individual magnification calibration tables so that we are all responsible for own measurements!

Other Useful Features on the TEM:

Although this is not a PC-driven microscope, the Hitachi-7100 TEM does have a range of microprocessor-controlled options which may be accessed via the cathode ray tube (CRT) screen at left of the EM column and a numbered and arrowed keypad set into the (extreme left) desk area of the EM. The most useful and routinely-used functions are as follows:

(1) Specimen Pre-Irradiation: This is a useful thing to do before you start examining your sections, especially if you are going to be acquiring images. Basically the programme scans the electron beam around the specimen to pre-heat it; in this way subsequent thermal drift in the EM should be minimized. 
Simply key in 20 on the numbered pad and press the INDEX button. 
Move the cursor on the CRT over the repeat number and select, say, 3 for a quick pre-heat or 10 for a thorough pre-heat (each scan takes c.50sec). 
Now move the cursor to START and press the DATA button (grey rectangular button on right). Wait for the pre-irradiation to finish before beginning to examine your grids!

(2) Memorizing Specimen Positions: I would recommend that you use this feature every time you use the EM. It will save you a great deal of time and avoid wasting imaging time. 
Key in 2 and press INDEX. 
Move the cursor over MEMORY POINT or DRIVE POINT (this toggles between the two by pressing the DATA button) and select MEMORY POINT. 
Now press select 00 and press the CLEAR button to remove previously stored memory points.
As you move around the sample and find interesting areas simply select MEMORY POINT and move the cursor over (P0), then select a number (up to 19 positions can be memorized) and press the DATA button twice to memories the position. 
To go back to a memorized position, move the cursor over MEMORY POSITION and toggle to DRIVE POSITION by pressing the DATA button. 
Now move the cursor over (P0), select a number and press the DATA button twice to drive to that position selected.

The specimen trace is another useful option (this is in the same screen as memorizing positions). Move the cursor over TRACE and toggle to ON by pressing the DATA button. If you are going to systematically scan a section it is a good idea to got to LOW MAG mode and move around the selected section to outline it; then when you got to higher mags (ZOOM function) you can scan left to right and down, etc. and see where you have been over the section.

(3) Image Rotation: This facility is available up to a magnification of 50,000X and is a very useful way to frame your image optimally. Simply rotate the IMAGE ROTATION knob just below the CRT. N.B. Remember to focus your image after you have rotated the image; rotation may alter focus and/or astigmatism slightly. 

You should now be able to achieve good final images which can be stored  digitally

TEM Troubleshooting (for camera troubleshooting, click here)

N.B. In all cases of problems with the TEM, please ask me for help if I am around. If I am not around at the time, and you cannot rectify the problem by reading the below, please mail me with a description of the problem.

    (1)    Losing the beam/image:

Firstly, check that the voltage to the electron gun has not been lost. The 'Acc voltage' button on the left console of the TEM should be illuminated. If it is not, press the 'acc voltage' ready/off button (just below the voltage buttons left console of EM) and then select the voltage required (usually 100kV). Now turn up the filament current (just below acc voltage button) as usual, i.e. quite slowly and fully clockwise. The electron beam should now be on.
Second Look at the 'HV Beam' current monitor on the upper left console of the TEM. The 'standing current' (i.e. with the filament not turned up via the filament control being turned fully clockwise) is 100. When you turn the filament up, this should increase to c. 125 on the HV Beam monitor. If this does not happen, and the voltage is on (see above), then the filament needs replacing.
If this happens whilst changing your grids or changing the objective lens aperture, for example, firstly make sure that it is simply not a problem with aperture alignment or grid positioning (over a grid bar or at the end of a specimen shift). The best thing to do initially is to go to 'low mag' mode and remove the objective lens aperture and see if you can image the whole grid.
However, if this happens suddenly as you are viewing your specimen, it may well be due to an intermittent alignment fault with the TEM. If so, please do the following: PLEASE DON'T try to twiddle lots of knobs to find the beam again. Instead, turn down the filament current and turn off the accelerating voltage.  Now press the RESET button below the CRT screen on the TEM left upper console. Turn on the accelerating voltage and turn up the filament current and you should find that everything is fine again. 
If this latter does not work, please do the following: Locate the 'Gun Tilt' (NOT TRANS) X and Y buttons that you can access behind the swing down panel lowermost left of TEM console (where the astigmatism correction knobs are located). Go to 'low mag' mode and remove the objective lens aperture and also remove the specimen rod to the 'park' position (pull out and then right). Now turn the X button fully counterclockwise, then sweep the Y button fully and slowly from left to right (or right to left), whilst constantly viewing the fluorescent screen (with all the room lights and PC monitor off, door closed as well is best). Then turn the X button a few mm clockwise and sweep through the Y button again. Keep on doing this until you should eventually see the beam. Once found, adjust both the X and Y buttons to acquire the brightest beam. Reintroduce your specimen and all should be well!

    (2)    Beam appears dimmer than usual

If the beam appears dim on first viewing, or suddenly dims as you are viewing, it may well be that the filament is ageing and has thus become misaligned. To align the filament, adjust the 'Gun Tilt' X and Y buttons to and fro in turn until you get the brightest beam possible. The Gun Tilt buttons are located behind the swing down panel lowermost left of TEM console (where the astigmatism correction knobs are located).
Otherwise, check that the objective lens aperture is properly aligned and that the beam is centred, of course, via the 'brightness centring' knobs on the right console of the TEM.

      (3)    Loss of movement in the X and Y specimen shifts of the TEM

If you find that the X and Y specimen shifts of the TEM suddenly stop working, this can usually be rectified by pushing in the recessed 'reset' button (with the wooden stick or the straight [not the delicate curved!!] forceps) that is located just to the left of the CRT screen displaying the image position, etc. (and just below the X and Y position numbers) on the upper left panel of the TEM as you are sat at it.